Features & Compatibility
r25c Diagnostic Audio Tymp Price in Bangladesh
DIMENSIONS AND WEIGHT
• L x W x H: 370 x 290 x 180 mm
• Net weight: 3.5 kg
TEST TYPES
• Tympanometry: Automatic Tympanometry, Acoustic
Reflex, Reflex Decay, Quick test: check and screening,
Multifrequency Timpanometry*
• Audiometry: Pure Tone test, Autothreshold, ABLB, Speech
Test, Stenger, GAP, DLF
USER INTERFACE
• Multilingual
PRINTER
• Built-in fast thermal printer with paper width: 112 mm
supplied as standard part
REPORTS
• Printed on thermal printer
• .pdf report created directly from the device and stored
on USB Pen drive with possibility to add patient data and
tests comments via the USB Keyboard (optional)
• Data transfer to PC using Resonance Management Data
Suite
COMMUNICATION PORT
• Nr.1 USB host type A • Nr.1 USB slave type B
ACOUSTIC REFLEX TESTS
• Reflex test method: Threshold , Automatic, Manual
• Stimulus duration: 0.5 , 1 or 2 sec.
• User selectable protocols for all test methods
• Ipsi or Contralateral stimulation for all reflex test
• Automated peak search functions available for all test methods
• Manual Reflex: manual control of all stimuli
• Reflex Decay: Threshold , Automatic, Manual
Stimulus duration: 10 or 20 sec.
FREQUENCIES AND INTENSITY RANGES
Ipsilateral:
• Pure tone level range (dB HL) from 50 to 110
• Noise level range (dB SPL) from 50 to 100
• Frequency: 500, 1000, 2000, 3000, 4000 Hz
• Noise: BBN, HP or LP
Contralateral:
• Pure tone level range (dB HL) from 50 to 120
• Noise level range (dB SPL) from 50 to 115
• Frequency: 250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz
• Noise: BBN, HP or LP • Frequency accuracy: ±1%
• Calibration accuracy ±3 dB • Level steps: 1, 2, 5 or 10 dB
• On/Off ratio: 70 dB minimum
Customer Reviews
Be the first to review “Resonance r25c Diagnostic Audio Tymp Price in Bangladesh” Cancel reply
Related Products
Natus Aurical® – Hearing Aid Fitting System
Aurical® streamlines hearing aid fitting for both you and your patients. This modular system offers real-ear measurements, verification tools, and a dedicated audiometer, all working seamlessly with Otosuite software. Achieve optimal hearing aid performance and personalized fittings with Aurical®, creating a smoother and more successful hearing care journey.
Natus Madsen A450 – PC-based Audiometer
The Madsen A450 streamlines hearing care with efficient audiometry and client counseling. This PC-based system simplifies air, bone, and speech testing with its user-friendly Otosuite software. Engage patients with counseling tools like hearing loss simulations and personalized reports. The Madsen A450 helps you deliver exceptional hearing care experiences.
Natus Madsen® Zodiac – with Diagnostic and Quick Check Gen 2 probes
Resonance r15c Screening Audio Tymp Price in Bangladesh
Resonance r14o OAE Screener Price in Bangladesh
Resonance r37a Clinical Audiometer Price in Bangladesh
Products Category
Contact Us
- 01859 244144
- hearingstorebd@gmail.com
- 152/2A-2, Rowshan Tower, (2nd Floor), Green Road Signal, Panthpath, Dhaka-1205, Bangladesh
Reviews
There are no reviews yet.