Inventis Clarinet Clinical Tympanometer
Equipments Brands |
Inventis |
---|---|
Equipments Categories |
Combined Audio – Tymp |
Model |
Clarinet |
Origin |
Italy |
Warranty |
2 Years |
Guaranteed Safe Checkout
Features & Compatibility
THE PROBE
Clarinet features an extremely light, small and sturdy metal probe, which can be very rapidly disassembled and cleaned. The probe is connected to the small Control Box, through which you can select the ear and start the exam.
ENDLESS FLOW PUMP
The speed of the pump is incredible: the sweep speed can reach 600 daPa/sec, without compromising on accuracy and precision of tympanogram plots. The pump features an endless airflow technology, improving the execution of the test when the seal is not optimal.
NOAH & MAESTRO INTEGRATION
Simply connect Clarinet to a USB port on your computer (no driver needed), and you can transfer exam results into your Maestro or Noah database. With the live view feature, you can have a wide view of the examination in progress on the computer display.
PROBE TONE | Frequency:
Level:
|
---|---|
ADMITTANCE MEASUREMENT SYSTEM | Measurement range:
|
PRESSURE SYSTEM | Rotary type pump with endless flow |
AVAILABLE TESTS |
|
TYMPANOMETRY | Viewing mode: absolute or compensated Measured components: Admittance [Y]; Susceptance [B] (Clarinet Plus only); Conductance [G] (Clarinet Plus only) Single or multiple scan (up to 3) Selectable sweep direction User selectable pressure range Maximum range: -600 to +400 daPa Safety limits: -800 daPa and +600 daPa Available rates: 15, 50, 100, 200, 300, 400, 600 daPa/sec and AUTO (200- 600 daPa/sec) Manual pressure adjustment Auto start feature |
ACOUSTIC REFLEX TEST | Ipsilateral and contralateral stimulation Non-acoustic stimulation Stimulus ON-Time: selectable from 0.5s to 2.5s Stimulus OFF-Time: selectable from 0.5s to 2.5s Four different tests available:
|
ACOUSTIC REFLEX DECAY TEST | Ipsilateral and contralateral stimulation Non-acoustic stimulation Test duration: 10 sec or 20 sec (selectable) |
ACOUSTIC REFLEX LATENCY TEST | Ipsilateral and contralateral stimulation Stimulus ON-Time: selectable from 0.3s to 1.7s Stimulus OFF-Time: selectable from 0.3s to 1.7s ON-Rate and OFF-Rate measurement: rate of the acoustic reflex slopes |
EUSTACHIAN TUBE FUNCTION TEST | ETF with TM intact: Williams test ETF with perforated TM: Toynbee test, with customable duration and pressure range |
QUICK TESTS | Two user programmable Quick tests available Each test consists of a sequence of
|
PRINTER | Optional integrated thermal printer Paper size: 112 mm |
CALIBRATION | Validity: 12 months All the parameters set through the device software |
COMPUTER INTERFACE | Connection: USB (driverless) Compatible software: Inventis Maestro |
DISPLAY | Type: Graphical color TFT LCD Size: diagonal 7”, 150 mm x 90 mm Resolution: 800 x 480 Resistive touch screen |
POWER SUPPLY | Without integrated printer: Main Unit: 6V, 1.67A dc. External Adapter: Medical grade 100-240 Vac 50/60 Hz 0.9-0.34AWith integrated printer: Main Unit: 6V, 3,6A dc. External Adapter: Medical grade 100-240 Vac 50/60 Hz 0.9-0.34A |
MECHANICS | Without integrated printer: Size (WxDxH): 32x32x15 cm / 12.6×12.6×5.9 in Weight: 2.0 kg / 4.4 lbsWith integrated printer: Size (WxDxH): 32x39x15 cm / 12.6×15.4×5.9 in Weight: 2.5 Kg / 5.5lbs |
FREIGHT PACKING | Size (WxDxH): 47x42x35 cm / 18.5×15.8×13.8 in Gross weight (without printer): 4.4 Kg / 9.7lbs Gross weight (with printer): 4.9 Kg / 10.8lbs |
Customer Reviews
Be the first to review “Inventis Clarinet Clinical Tympanometer” Cancel reply
Related Products
Natus ALGO® 7i Newborn Hearing Screener
The ALGO® 7i ensures early detection of hearing loss in newborns. This portable device uses proven AABR technology for accurate screening. Its user-friendly design with on-screen guidance simplifies testing for healthcare professionals. The ALGO® 7i empowers fast, reliable results, giving newborns the best chance for timely intervention and healthy development.
Natus Madsen® Accuscreen® – Newborn Hearing Screener OAE/ABR
The Madsen® Accuscreen® offers a two-step approach to newborn hearing screening, combining OAE (Otoacoustic Emissions) and ABR (Auditory Brainstem Response) tests. This comprehensive approach delivers accurate results while improving efficiency. The Madsen® Accuscreen®'s portability and user-friendly design streamline the screening process for healthcare providers. Early detection of hearing loss allows for prompt intervention, giving newborns the chance for optimal development.
Natus Madsen® Zodiac – with Diagnostic and Quick Check Gen 2 probes
Natus Otoscan® – 3D Ear Scanning Solution
Otoscan® offers a revolutionary upgrade to hearing care with its 3D ear scanning technology. Ditch the uncomfortable, messy traditional impressions and get a faster, more precise fit for custom hearing aids and earmolds. Otoscan's digital scans capture intricate details, leading to increased comfort and reduced remakes. This innovative solution enhances patient experience and streamlines workflow for hearing professionals.
Path Medical Sentiero Advanced (OAE, ABR, ASSR device)
Universal Galaxy I Audiometer
The Universal Galaxy I Audiometer facilitates comprehensive hearing evaluations. This advanced system allows for air and bone conduction testing, providing valuable data for accurate diagnoses. Its user-friendly design (if information available) ensures a smooth testing experience for both patients and professionals. The Universal Galaxy I Audiometer empowers informed decision-making for better hearing healthcare.
Products Category
Quick Links
152/2A-2, Rowshan Tower,(2nd Floor), Green Road Signal,Panthpath, Dhaka-1205, Bangladesh
© 2024 . All Rights Reserved by Hearing Store
Reviews
There are no reviews yet.